X – Ray Diffraction Laser THERMO LOGG Contact Angle Analyzer Langmuir – Blodgett Film Deposition Scanning Electron Microscope with EDS (X-ray spectrometry) Small Angle X-Ray Scattering Apparatus Wide Angle X-Ray Scattering Apparatus Mercury Porosimeter Mass Spectrometer Nitrogen Porosimeter ultra-microtome AA GC-MS Scanning Electron Microscope with EDS (X-ray spectrometry) Proteome analysis [Proteomics] Remote Measurement System Transmission Electron Microscope CNC ΑGIECharmilles ΑCTSPARK FW-1P [CNC AGIE] CNC DMG CTX 510 Eco PHOTRON FASTACAM SA3 INSTRON 8801 Testing Device ROMER OMEGA R-SCAN & 3D RESHAPER LASER Cutter Pantograph with extra PLASMA torch CNC ΙDA XL 1200 Optical and Contact Coordinate Measuring Machine TESA MICRO-HITE 3D  RSV-150 Remote Sensing Vibrometer Ground Penetration Radar [GPR] Audio Magneto Telluric Optical Time Domain Reflectometers [OTDR] Non ion Rad Electric e-mat analysis Thermogravimetric Analyzers - Differential Scanning Calorimetry Magnetron Deposition Metal Deposition Grid Computing Center

Instrument for measuring the thermal conductivity factor [FOX200/300]

The FOX200 instrument can be used to measure apparent thermal conductivity of thermal insulation materials. It was designed according to the ASTM C518-04 “Standard Test Method for Steady – State Thermal Transmission Properties by Means of the Heat Flow Meter Apparatus”.       

Contact person:

Eleni Apostolidou
tel. (+30) 2510462225
This email address is being protected from spambots. You need JavaScript enabled to view it.

FOX200-300

 

The instrument is calibrated using the NIST SRM 1450b (Standard Reference Material of the National Institute of Standards and Technology, and the special high accuracy (0.5%) EPS (Expanded Polystyrene) NIST Standard. The calibrations were done with a 25 oC ή 20 oC temperature difference and are saved in instrument’s memory.

 

Specifications:

Thermal conductivity range                              0.01 – 0.02 W m-1 K-1

Maximum temperature of hot plate                    75 oC

Minimum temperature of cold plate                   -20 oC

Temperature control stability                            ±0.03 oC

Thickness measurement accuracy                     ±0.025mm

Maximum sample size                                      203 x 203mm

Minimum sample size                                       100 x 100mm

Maximum sample thickness                               51mm

Interaction with pc and printer                          RS-232 serial interface

 

Wintherm 32 Software for connection the FOX200 to a PC

The software can manage the operation of the instrument from a PC, record the results in a file and can be programmed to execute experiments in a big range of temperatures for the same sample.